Abstract

The design and performance characteristics of a new multielement flameless atomic absorption spectrometer are presented. Radiation from four hollow cathode lamps is combined by beam splitters and directed through a Varian model 63 carbon rod atomizer and then into a special monochromator with a separate exit slit for each element. The light exiting from all slits is directed to a single photomultiplier tube. Absorbance signals from different elements are distinguished with a time multiplex approach. Data acquisition and presentation is accomplished with a PDP-11 computer system. The absolute light levels are lower than those normally encountered with single element atomic absorption resulting in signal shot noise limited measurements. Detection limits (in ppb) of 0.2 (Cd), 2 (Mn), and 8 (Pb) were determined under multielement conditions. The measurement precision of moderate absorbance multielement solutions was determined to be 7% with manual injection of 5 μl samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.