We present a new design for mounting a cryogenically cooled Si crystal which gives greatly improved beam stability. The design has been successfully implemented at the University of Michigan, Howard University, Bell Laboratories-Lucent Technologies Collaborative Access Team (MHATT-CAT) 7ID Beamline of the Advanced Photon Source. Before the installation of the new crystal mount, our Si (lll) cryogenically cooled monochromator was sensitive to the pressure fluctuations of the liquid nitrogen coolant, such that the angle of incidence on the first crystal varied linearly with the applied pressure in the cooling lines, causing beam motion of about 250 μm, 60 m from the source. The key element of the design is a symmetrically positioned cooling manifold which balances the forces caused by pressure fluctuations. With this new mount, the typical beam stability is now about 10 μm, comparable to the source stability.