Abstract

A 2-D multi-wire position sensitive detector for X-ray diffraction and small angle X-ray scattering studies is described. The detector has an active area of 100 mm × 100 mm and consists of an anode plane with 10 µm SS wires at 3 mm spacing and a pair of orthogonal cathode readout planes with 25 µm SS wires placed at 1.5 mm spacing. The position information is obtained using charge division method and recorded using a laboratory built data acquisition system. The resolution and gas gain was measured for 5.9 keV X-rays (55Fe-source) as a function of the anode wire voltage and gas pressure. It was observed that the proportional region of the PSD at 100 kPa pressure extended up to a high voltage value of around 1.5 kV and it shifted to high values up to 2 kV for gas pressure of 300 kPa. The energy resolution improved from 18% (FWHM) to 12% with increase in pressure. The spatial resolution of the PSD also showed improvement, with a value of 1.2 mm × 1.4 mm at 300 kPa gas pressure. A maximum gain of 5 × 104 is obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.