Low cost coaxial surface junction thermocouples (CSJTs') have been fabricated in-house and calibrated to measure the transient surface temperature rise within a UNITEN's shock tube wall facility, consisting of K-type coaxial thermocouple elements. The aim of this paper is to explain the design technique of the CSJTs' and the difficulties that have occurred during the fabrication process. The microstructural analysis and the chemical characterization for these types of thermocouples have also been carried out to verify the surface morphology and to qualitatively evaluate the CSJT materials composition. The preliminary testing was performed to demonstrate the performance of these thermocouples to be used for measuring the surface temperatures and heat transfer rates under transient conditions. The preliminary results from shock tube tests have shown that these thermocouples have a time response on the order of microseconds and were suitable for making heat transfer measurements in highly transient conditions. It was concluded that the current construction technique produced gauges that were reliable, reproducible, rugged and inexpensive.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.