Abstract

This paper presents a new design and fabrication process of a multiple-thickness electrochemical cantilever sensor, in order to assess the role of the cantilever’s thickness on the chemically-induced mechanical effects. Each cantilever can act not only as a functionalized cantilever, but also as an independent working electrode (WE) for electrochemical measurement. The different thicknesses of the silicon nitride layer are achieved by successive masking and reactive ion etching of partially overlapping openings at a low etch rate (10.8nm/min, 15.8nm/min, 20.1nm/min, or 26.5nm/min). A small-scale thickness difference (<30nm) is successfully obtained. One advantage of this fabrication process is that the thickness distribution of cantilevers can be altered and broadened by combination of different RIE recipes or modification of the etching time. In addition, the integration of the cantilever chip with a fluidic cell, a printed circuit board (PCB) and a temperature-controlled plate to form a hybrid system is also addressed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.