Abstract

A biosensor for bacterial detection was developed based on microelectromechanical systems, heterobifunctional crosslinkers and immobilized antibodies. The sensor detected the change in impedance caused by the presence of bacteria immobilized on interdigitated gold electrodes and was fabricated from (100) silicon with a 2-/spl mu/m layer of thermal oxide as an insulating layer. The sensor active area is 9.6 mm/sup 2/ and consists of two interdigital gold electrode arrays measuring 0.8 /spl times/ 6 mm. Escherichia coli specific antibodies were immobilized to the oxide between the electrodes to create a biological sensing surface. The impedance across the interdigital electrodes was measured after immersing the biosensor in solution. Bacteria cells present in the sample solution attached to the antibodies and became tethered to the electrode array, thereby causing a change in measured impedance. The biosensor was able to discriminate between different cellular concentrations from 10/sup 5/ to 10/sup 7/ CFU/mL in pure culture. The sample testing process, including data acquisition, required 5 min. The design, fabrication, and testing of the biosensor is discussed along with the implications of these findings toward further biosensor development.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.