Abstract

In this paper, the design and performance of a collimated Knudsen source, which has the benefit of a simple design over recirculating sources, is discussed. Measurements of the flux, transverse velocity distribution, and brightness of the resulting rubidium beam at different source temperatures were conducted to evaluate the performance. The scaling of the flux and brightness with the source temperature follows the theoretical predictions. The transverse velocity distribution in the transparent operation regime also agrees with the simulated data. The source was tested up to a temperature of 433 K and was able to produce a flux in excess of 10(13) s(-1).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.