Abstract

Flow stability of samplers is a key factor in air sampling. However, most of the existing air samplers have the problem of poor stability. In order to enhance the stability, a constant flow air sampler (HL-1) has been designed: a single chip processor was added to the flow control part. The performance of the sampler was presented and comparison with a conventional sampler was made. Inter-sampler comparison tests were carried out by comparing (a) the indicating error (b) the repeatability (c) the consistency of the flow and (d) the timing error with a traditional sampler. The results showed the improved sampling apparatus performed better than the traditional one in respect of stability. It could operate in a predetermined range of flow and volume in an entirely reliable and predictable manner.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.