Abstract

Nursing is a physically demanding job characterized by a high prevalence of fatigue and musculoskeletal disorders. One of the high-exertion and repetitive nursing tasks is the manual connection of an infusion set to a medical fluid bottle. Such physical work can be eased by the design of new hand tools. Correspondingly, this study designed and ergonomically assessed an infusion set connector tool (ISCT) and compared it with that of manual connection. First, a prototype of ISCT was designed to perform infusion set connecting task in the mechanical form. Subsequently, 12 nurses were asked to connect an infusion set to medical bottle in the form of manual and mechanical tasks and these tasks were evaluated using ergonomic indices including muscular activity level, force, posture, and subjective (Borg scale CR10) measures. Results showed that the activity levels (root mean square) of the extensor digitorum communis, flexor carpi radialis, biceps, triceps, and deltoid muscles remarkably decreased when the nurses used the ISCT. The postures of the wrist, arm, and shoulder regions were corrected from Rapid Upper Limb Assessment action level 3 to 2 when the designed tool was used. Additionally, the subjective perception of exertion was significantly lower with the use of the prototype.