In this work, FLUKA design of a beam shaping assembly (BSA) has been carried out for sealed D–T neutron generator to have application in BNCT (boron neutron capture therapy). Out of several materials, the design has been optimized for tungsten multiplier, D2O moderator, TiF3 fast neutron filter, cadmium thermal neutron filter and bismuth gamma filter surrounded by lead reflector. The designed BSA would provide appropriate beam intensity and quality at lower source strength. The dosimetric evaluation of BSA carried out with Snyder head phantom is found to be satisfactory for treatment of deeply seated tumor in 115 min. The proposed design of BSA for compact D–T neutron generator can be installed in hospital premises for BNCT treatment with better social acceptance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.