Abstract

The design and calibration of the dispersive device in a hyperspectral imager significantly affect the performance of hyperspectral imaging, especially the spectral accuracy. To achieve high-accuracy hyperspectral imaging over the visible band, firstly, the geometric and dispersive parameters of the double Amici prism (DAP) that serves as a dispersive device in the direct-vision push-broom compressive hyperspectral imager (PBCHI) are designed and optimized; secondly, a calibration method based on the numerical calculation of the DAP model is put forward, which can turn the conventional pixel-wise dispersive shift calibration by a monochromator into a group of numerical calculations; lastly, a PBCHI prototype is built to test the performances of the designed and calibrated DAP and the hyperspectral imager. The calibration experiments demonstrate that the mean squared error (MSE) of the dispersive pixel shifts calibrated by the proposed numerical method is 0.1774, which indicates the calibration result of the proposed method is consistent with the directly calibrated result. Furthermore, after this numerical calculation, the spectral signatures of the reconstructed cubes of the DAP-based PBCHI system show consistency with the ground truth. This work will benefit the design and calibration of the DAP-based hyperspectral imager.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call