An instrument is described that permits fluorometric titrations in a large, open compartment that accommodates titration vessels of a wide variety of shapes and sizes. The influence of ambient light is excluded by applying geometric exclusion in combination with electronic monitoring and compensation. In test titrations of copper and calcium with EDTA and calcein, the instrument performed satisfactorily.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.