Abstract
To fulfill the need for reliable and consistent medical training of the neurological examination technique to assess ankle clonus, a series elastic actuator (SEA) based haptic training simulator was proposed and developed. The simulator's mechanism (a hybrid of belt and linkage drive) and controller (impedance control) were designed to render a realistic and safe training environment. Benchtop tests demonstrated that the prototype simulator was able to accurately estimate the interaction torque from the trainee (average RMSE of 0.2 Nm) and closely track a chirp torque command up to 10 Hz (average RMSE of <0.22 Nm). The high-level impedance controller could switch between different clinically encountered states (i.e., no clonus, unsustained clonus, and sustained clonus) based on trainee's assessment technique. The simulator was evaluated by a group of 17 experienced physicians and physical therapists. Subjects were instructed to induce sustained clonus using their normal technique. The simulator was assessed in two common clinical positions (seated and supine). Subjects scored simulation realism on a variety of control features. To expedite controller design iteration, feedback from Day 1 was used to modify simulation parameters prior to testing on Day 2 with a new subject group. On average, all subjects could successfully trigger a sustained clonus response within 4-5 attempts in the first position and 2-3 in the second. Feedback on the fidelity of simulation realism improved between Day 1 and Day 2. Results suggest that this SEA-based simulator could be a viable training tool for healthcare trainees learning to assess ankle clonus.
Highlights
In the scope of this work, we focused on developing a simulator to help train healthcare learners to perform ankle clonus assessment, a common task during neurological examinations (Fig. 1)
Our goal was to design a torque-controlled haptic device that rendered a realistic feeling of the muscle response of a patient with ankle clonus to trainees
The torque estimated by the deflection of the series springs matched well with the torque sensor reading, where the average root mean square error (RMSE) was 0.20 Nm (Fig. 6)
Summary
D URING neurological examinations, clinicians need to perform manual assessment techniques (e.g., passive stretching of affected muscles, tapping the tendon) to elicit clinical signs to diagnose neurological conditions and rely on haptic experiential knowledge to evaluate muscle tone. Instructors usually bring in a small number of practice patients into the classroom or let students pretend to be the patient for each other, which leads to limited practice opportunities and inconsistent training outcomes [7]. This training challenge calls for more accessible and consistent ways to provide training opportunities for students that replace or reduce the need for human practice patients [8]. One promising approach is the deployment of medical training simulators
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.