Abstract

A multi-frequency bioimpedance measurement prototype is proposed, validated and characterized. It consists of an Improved Howland Current Source controlled by voltage, a load voltage sensing scheme through a discrete 3-opamp instrumentation amplifier, a phase and quadrature demodulation setup through analog multipliers, and digitization and processing of the signals using a digital benchtop multimeter. The electrical characterization of the measurement channel was done for resistive loads only, on four different circuits. Measurements were made on 10 frequencies, from 100 kHz to 1 MHz, with 10 load resistances, from 100 Ω to 1 kΩ, to obtain linearity, absolute error and frequency response. The best performance among the four circuits was a maximum absolute error of 5.55 %, and −1.93 % of load current variation at the worst case scenario.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.