Abstract

Design and characterization of a general-purpose spectrometer for recording time-resolved emission spectra of typical fluorescent species is described. The system is based on a high repetition rate amplified Ti : sapphire system, an optical Kerr shutter for gating the emission, and a polychromator plus charge-coupled device (CCD) detection system. Using 1 mm of liquid benzene as the Kerr medium, and optics designed to provide high polarization quality, emission spectra of dilute solutions of solutes with nanosecond lifetimes can be recorded with good signal-to-noise ratios. The current spectrometer uses excitation wavelengths near 390 nm and provides spectra over the wavelength range 400-650 nm with 4 nm resolution and instrument response times of 450 fs (full width at half-maximum, FWHM). Selected applications are described to demonstrate the utility of this instrument.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.