Abstract

Abstract Electrical Impedance Tomography is used to image the cross-sectional conductivity of an object. It is clinically used for high-frame-rate imaging of lung ventilation. Most current systems use very limited current injection waveforms and patterns. We have developed a flexible system for researching alternatives. This paper covers our voltage conditioning module’s design and characterisation. The device was shown to work up to the desired 10 MHz. The responses were uniform with minor gain mismatch between all channels after manufacturing variations. The low-frequency (<1 MHz) crosstalk is on the order -55 dB, and the worst cases around 8 MHz are around -35 dB.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.