Abstract
This paper describes the design and characterization of a charged particle imaging system composed of a position sensitive detector and residual range detector. The position detector consists of two identical overlying and orthogonal planes each of which consists of two layers of pre-aligned and juxtaposed scintillating fibres. The 500μm square section fibres are optically coupled to two Silicon Photomultiplier arrays using a channel reduction system patented by the Istituto Nazionale di Fisica Nucleare. The residual range detector consists of sixty parallel layers of the same fibres used in the position detector each of which is optically coupled to a Silicon Photomultiplier array by wavelength shifting fibres. The sensitive area of the two detectors is 9×9cm2. Characterising the position sensitive and the residual range detectors to reconstruct the radiography, is fundamental to validating the detectors’ designs. The proton radiography of a calibrated target in imaging conditions is presented. The spatial resolution of the position sensitive detector is about 150μm and the range resolution is about 170μm. The performance of the prototypes were tested at CATANA proton therapy facility (Laboratori Nazionali del Sud, INFN, Catania) with energy up to 58MeV and rate of about 106 particles per second. The comparison between the simulations and measurements confirms the validity of this system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.