Abstract

Two fingers strength is an indicative measurement of pinch impairment. Conventionally, Fugl Meyer Upper Extremity Assessment (FMA-UE) is the primary standard to measure pinch strength of post-stroke survivors. In literature, the evaluation method performed by the therapist is subjective and exposed to inter-rater and intra-rater reliabilities. Recently, force-sensing resistors were implemented to measure two fingers force, but these sensors are subjected to nonlinearity, high hysteresis, and voltage drift. This paper presents a design of pinch force measurement based on the strain gauge. The pinch sensor was calibrated within a range of between 0 N to 50 N over a pinching length of 20 mm with a linearity error of 0.0123% and hysteresis of 0.513% during the loading and unloading process. The voltage drift has an average of 0.24% over 20 minutes. The pinch force measurement system reveals an objective pinch force measurements in evaluating the rehabilitation progress of post-stroke patients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.