Abstract

An optical sensing system has been developed using a pair of orthogonally placed position sensitive detectors (PSD) to track 3D displacement of a microsurgical instrument tip in real-time. An infrared (IR) diode is used to illuminate the workspace. A ball is attached to the tip of an intraocular shaft to reflect IR rays onto the PSDs. Instrument tip position is then calculated from the centroid positions of reflected IR light on the respective PSDs. The system can be used to assess the accuracy of hand-held microsurgical instruments and operator performance in micromanipulation tasks, such as microsurgeries. In order to eliminate inherent nonlinearity of the PSDs and lenses, calibration is performed using a feedforward neural network. After calibration, percentage RMS error is reduced from about 5.46 % to about 0.16%. The system RMS noise is about 0.7 mum. The sampling rate of the system is 250 Hz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.