Abstract
The design and construction of a Full-Stokes Imaging polarimeter is presented. The device uses a telescope optical system and a polarization state analyzer (PSA) to obtain polarimetric images on a CCD. The PSA employs two liquid crystal variable retarders (LCVR) and a linear polarizer to measure the four Stokes parameters. The Stokes polarimetry method used in this paper is based on the application of six combinations of retardance values on the LCVRs. A well-known method is used to extract all the Stokes vector parameters from this intensity data. Due to experimental errors, a calibration is necessary. The calibration method used in this paper, also calculates the errors in the experimental set-up by fitting the experimental intensity measurements for the calibration samples to a theoretical polarimeter with errors. In this case, we used incident 45° polarized light to control the output polarization, and six calibration samples. The errors calculated in the method include the axes alignment errors and the errors in the retardance values of both LCVRs. The acquisition of Stokes images used a telescope optical system with a CCD camera.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.