Abstract

The ventilator is a supporter of respiratory needs which is very important for the patient so that there are several parameters that must be monitored specifically, such as the measurement of pressure and flow rate used in the ventilator system, the accuracy of which must be in accordance with the accuracy of the respirator. One of the important parameters to monitor is PIF (Peak Inspiratory Flow) which is the peak inspiratory flow rate given through the ventilator. PIF that is too high or too low can cause adverse effects on the patient. PIF monitoring can be seen through the PIF value and waveform on the PIF. Monitoring the waveform of the PIF will be very useful to improve the results of using the ventilator. The purpose of this research is to get the accuracy and precision of the sensor to display the waveform of the ventilator output. The procedure carried out is to use the F1031V sensor to detect the flow generated by the ventilator and then detect the PIF value and PIF waveform. From this research, the measurement of accuracy and precision of the F1031V sensor to detect PIF and generate a waveform graph is said to be good. This is because the highest error value is ±2.04% at the 20 LPM setting. While the value of the largest standard deviation at the 30 LPM setting is 1.517 and the greatest uncertainty value at the 30 LPM setting is 0.061. Then, the largest correction value is found in the setting of 20 LPM and 30 LPM, namely 0.4. PIF monitoring is carried out to maximize patient care and reduce the breakdown time on the ventilator.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call