Abstract
In the present study, a novel subaxial cervical pedicle screw placement guide device was designed and developed. In cervical specimens (C3-C7), a pedicle screw was inserted into the left pedicle using the guide device with a keyhole partial laminectomy and tapping technique, and the right pedicle by drilling using the Abumi technique. After removing the pedicle screws, the channel wall of each pedicle screw was probed with a pedicle probe. The vertebral body was then dissociated for direct observation of the screw channel. Among the 10 specimens, 2 of the 50 pedicles (4%) in the guide device group were perforated. Screw placement failed in 8 of 50 pedicles (16%) in the Abumi technique group. Significant differences were observed in the outcomes for the guide device and Abumi technique groups. The subaxial cervical pedicle screw placement guide device developed in the present study decreased the failure rate of pedicle screw placement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.