Abstract
A new versatile phase separator for flow injection solvent extraction has been designed and its analytical performance investigated. Manifolds for various instrumentation and the corresponding operating procedures have been established. Experimental results show that the newly designed phase separator possesses several advantages over the existing ones, such as long term stability, robustness, sufficient phase separation and integral in situ collection of the solvent extracts at different phase flow rate ratios. The detection limits for cobalt, gold, iodide, iron, lead, nickel, nitrate and tungsten were 1.4, 0.45, 70, 1.9, 160, 0.9, 137 and 23 ng ml −1, and the relative standard deviations 2.5, 1.3, 2.2, 2.9, 4.8, 3.6, 1.4 and 3.3%, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.