Abstract

Magnetic resonance imaging (MRI) technique is widely used to capture the images of the liquid items inside the human body. The radio-frequency (RF) coil is one of the important modules present inside an MRI system, which plays a major role in image quality. In this work, a microstrip-based high-pass RF birdcage coil is proposed for 1.5 T MRI. The cylindrical-shaped birdcage coil consists of 12 microstrip radiating elements and tuning capacitors to achieve a resonance at 63.85 MHz. The coil is made up of 10 mm polytetrafluoroethylene substrate coated by a conducting transmission line of desired length and width. A finite difference time domain simulation is carried out to analyze the return loss (S11), magnetic field homogeneity and Specific Absorption Rate (SAR) parameters of the RF coil. The SAR values of the proposed microstrip-based 1.5 T birdcage coil was compared with 3 T RF birdcage coil. The simulation results indicate the proposed birdcage coil structure gives optimal values of S11, magnetic field homogeneity and SAR.