Abstract

A new tympanic thermometer is analyzed and testedexperimentally. An electrically calibrated pyroelectric detector ofspecial configuration is employed to determine a person's bodytemperature. An energy-storage, power-supply-isolated capacitor isused as the electrical heating reference. The new thermometerdesign has an accuracy within ?0.1 degrees C with a 90% confidence and isimmune to ambient temperature, detector aging, and parametervariations. An equivalent-circuit model is established in theanalysis to account for the heat exchanges among the tympanum, thesurroundings, and the detector as well as for the electrothermalbehavior of the detector. The model provides effective simulationof the thermometer with PSPICE. Critical parameters governing theaccuracy and the limitation of the tympanic thermometer are alsopointed out by the simulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.