Abstract

Surgical robot with a remote center of motion (RCM) plays an important role in minimally invasive surgery (MIS) field. To make the structure more compact and have higher rigidity and accuracy, a new type of parallel surgical robot with three degrees of freedom (DOFs) is developed. The detailed design of the mechanism is provided in the paper. To better study the characteristics of the mechanism, static analysis is proposed in the paper. Firstly, establish the equations of the force and torque through the analysis of the branch and the mobile platform. Next, obtain the expression of the active force and constraining force of the joints. One numerical example is given. In order to analyze the dynamic performance of the UPR-2UPRR system, the Kane method is used to build the dynamic model. The speed, acceleration, partial velocity and partial angular velocity are calculated. Then the dynamic equation is given after analyzing the generalized active force and inertial force. At last, the driving force of each telescopic rod is given.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.