Abstract
Ambient ionization mass spectrometry (MS) techniques, such as desorption electrospray ionization (DESI), have been increasingly used due to their simplicity, minimal sample preparation requirements, and potential applications in the field and the clinic. However, due to their intrinsic nature, the performance of these methods is susceptible to variations in ambient conditions. Here, we present data that suggests DESI-MS analysis becomes inconsistent below a relative humidity (RH) level of ~ 35%. At low RH, we hypothesize that the DESI spray is subjected to frequent electrical discharges, resulting in unstable ionization and atypical mass spectra. Consequentially, poor image quality is observed when used for tissue imaging. Our results suggest that RH control should be considered in DESI-MS experiments to assure data quality. Graphical Abstract ᅟ.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Society for Mass Spectrometry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.