Abstract

Electrical mobility analysis is the most efficient technique for measuring aerosol particle size distributions in the submicron size range. Recent advances in aerosol science underline the need of fast measurements of particle size spectra in this range, and therefore a great amount of effort has been focused towards this direction. This paper provides the description and a theoretical framework for the analysis of a fast-response differerntial mobility spectrometer (DMS). In common with other instruments of its category, it consists of a particle charger, a classification column, and a series of detectors. Passing the sample flow first through a corona-wire diffusion charger that sets a charge on the particles, the aerosol is introduced around the central rod of an inside-out cylindrical classifier equipped with a series of isolated electrode rings connected to sensitive electrometers. Current readings produced by deposition of the charged particles on the electrometer rings are then translated to partic...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.