Abstract
The objective of this paper is the development of a much needed facility to calibrate environmental radiation field measuring devices. A well characterised standard low-level radiation field facility can provide the basis for instrument calibrations and dispute resolution. The experimental facility which was installed underground in a large low radiation scatter room, will be described. The facility was approached using multidimensional Monte Carlo program (MCNP-Monte Carlo N-Particle code) which models the detector/source to estimate detector efficiencies. This work is the outcome of the close cooperation between the Ionizing Radiation Division of NIST, USA and Nuclear Engineering Laboratory of Democritus University of Thrace, Greece.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.