Abstract
Reactive dyes, such as Procion Yellow, Procion Red, and Procion Blue, were simultaneously determined in aqueous ternary solutions using derivative cathodic stripping voltammetry following adsorption on a hanging mercury drop electrode. An electrochemical cell containing Britton–Robinson buffer solution at pH 8 was used as the supporting electrolyte. The following experimental conditions were established: −0.100 V deposition potential, 80 s deposition time, 50 mV pulse amplitude, 50 mV s−1 scan rate, and 0.40 mm2 maximum mercury drop size. A linear response was observed over the 0.630 × 10−3 to 1.050, 1.441 × 10−3 to 1.572, and 0.198 to 1.570 mg l−1 ranges for Procion Yellow, Procion Red, and Procion Blue respectively. The detection limits were 0.210 × 10−3, 0.480 × 10−3, and 0.066 mg l−1 respectively, with a relative standard deviation (n = 3) of 1.22%. The accuracy of the method was evaluated by derivative spectrophotometry, certifying the results as a function of proximity. All results were similar, indicating that derivative cathodic stripping voltammetry may be efficiently applied for the simultaneous determination of textile dyes in aqueous ternary solutions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.