Abstract

A no-reference method to detect and suppress ringing effect in MRI images is suggested. The ringing detection method is based on finding the area where ringing effect is likely to appear and calculating the ratio of average edge-normal and edge-tangential derivatives moduli in this area. The area consists of pixels with the certain distance to basic edges--sharp edges that are distant from other edges. The proposed ringing suppression method is based on the projection onto the set of images with bounded total variation with ringing level control.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.