Abstract

Shot peening can increase the fatigue strength of commercially available surgical plates made of 1.4435 alloy by 40% even in a corrosive environment. Our investigations show that residual stresses resulting from shot peening are reduced by additional bending of the plates. In such plates smaller tensile residual stresses were found than after polishing of the plates. Bending of polished plates results in considerable tensile residual stresses. The hardening achieved by shot peening is not reduced by bending. As the fatigue strength of soft materials depends mainly on the hardening and less on the residual stresses, only little influence of the changed residual stresses on the fatigue strength can be expected. Shot peening of surgical implants thus means an improvement in quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.