Abstract

The depth scale calibration of a SIMS depth profile requires to determine the sputter rate used for the analysis from the crater depth measurement. An in situ crater depth measurement system based on the heterodyne laser interferometer has been developed. Experimental results demonstrate that crater depths can be measured from nanometers to micrometers range with an accuracy better than 5% in different matrices and a repeatability of 1%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.