Abstract

Secondary ion mass spectrometry analyses were carried out using a metal cluster complex ion of Ir4(CO)7+ as a primary ion beam. Depth resolution was evaluated as a function of primary ion species, energy, and incident angle. The depth resolution obtained using cluster ion bombardment was considerably better than that obtained by oxygen ion bombardment under the same experimental condition due to reduction of atomic mixing in the depth. The authors obtained a depth resolution of ∼1nm under 5keV, 45° condition. Depth resolution was degraded by ion-bombardment-induced surface roughness at 5keV with higher incident angles.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.