Abstract

Depth of interaction (DOI) decoding capability shows great application potential for positron emission tomography (PET), especially for high-resolution small-animal PET or long axial field of view PET systems. On the basis of GATE optical simulation results, we presented a novel $9\times 9$ lutetium–yttrium oxyorthosilicate array with 10-mm rectangular light-sharing windows (RLSWs) and assembly the detector with nine crystals coupled to one silicon photomultiplier (SiPM) method; this caused DOI-related position shifts in the crystal decoding spots of the flood map. Homogeneous radiation was used to establish transfer functions to convert spot shifts into DOI measurements. The conventional collimated radiation experiments were used to assess the accuracy of the DOI measurements. Flood map results show that 81 segmented crystals can be clearly resolved, including 16 crystals adjacent to two RLSW windows (M-type), 40 crystals adjacent to one RLSW window (E-type), and 25 crystals not adjacent to any RLSW window (C-type). The mean absolute error of all DOI-capable crystals, M-type crystals, and E-type crystals are 2.15 ± 0.61, 1.95 ± 0.50, and 2.23 ± 0.63 mm, respectively. The novel nine-crystals-to-one-SiPM coupling technology is a low cost approach for constructing high-performance DOI detector modules. Detectors using nine-crystals-to-one-SiPM coupling technology are good candidates for high-resolution preclinical PET and for total-body PET.