Abstract

Background and purpose: The normalisation depth for determination of output factors in photon fields has frequently been the depth of dose maximum. At high energies the contribution from contaminating electrons is significant at dose maximum and is critically dependent on the beam geometry parameters, which is why a larger depth should be preferred. Materials and methods: The effect of electron contamination was studied using a purging magnet to remove charged particles from the treatment head and a helium bag to minimise production between the head and the phantom. Results: A depth of 10 cm was found to be beyond the range of the contaminating electrons for photon energies up to 20 MV (TPR 10 20 = 0.772). However, at 50 MV (TPR 10 20 = 0.810) contaminating electrons contribute 2–3% to the absorbed dose at 10 cm depth. Conclusions: 10 cm is recommended as both reference and normalisation depth for all megavoltage photon beam qualities, i.e. 60Co and X-rays from accelerators up to 50 MV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.