Abstract

We demonstrate a method to select different layers in a sample using a low coherent gating approach combined with a stable common-path quantitative phase imaging microscopy setup. The depth-filtering technique allows us to suppress the negative effects generated by multiple interference patterns of overlaying optical interfaces in the sample. It maintains the compact and stable common-path setup, while enabling images with a high phase sensitivity and acquisition speed. We use a holographic microscope in reflective geometry with a non-tunable low coherence light source. First results of this technique are shown by imaging the hardware layer of a standard micro-controller through its thinned substrate.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call