Abstract

To compare the dependency of dose response of five CCD-based digital intra-oral radiographic systems on tube voltage. Characteristic curves for Sens-A-Ray (old and new generations; Regam Medical Systems, Sundsvall, Sweden), Computed Dental Radiography (Schick Technologies, Long Island City, NY, USA), CompuRay (Yoshida, Tokyo, Japan), and VIXA (Dentsply/Gendex, Des Plaines, IL, USA) were generated between 60 and 90 kVp at intervals of 10 kVp and corrected by subtraction of dark current. Their sensitivities were compared. The sensitivity of the old generation Sens-A-Ray decreased with increase in tube voltage whereas in the other systems it increased. In the newer systems using a rare earth intensifying screen as a scintillator, the sensitivity increases with increase in tube voltage, thus reducing the radiation dose to the patient.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.