Abstract

The position dependency of gas amplification in the proportional counter (PC) is investigated. We have been developing one-dimensional position sensitive PCs for MAXI/GSC and HETE/WXM and found that anomalous gas amplification occurs in a high bias voltage, even while the PC is still operated in the proportional region. This effect depends on the position where the X-ray is absorbed. Therefore it appears as a hard tail, a soft tail, or a broad peak in the traditional PC, depending on the shape of the gain curve across the cell. It degrades the apparent energy resolution. Especially, a position sensitive proportional counter (PSPC) is operated with rather high bias voltage to give higher positional resolution. We encounter the difficulty to achieve good position and energy resolutions at the same time. In this work, we have examined the anomalous gas amplification for various gas mixtures of Xe + CO<SUB>2</SUB>, Ar + CO<SUB>2</SUB> and Ar + CH<SUB>4</SUB>, for gas gain up to approximately 20000, and for energies from 6 to 17 keV to understand the phenomena.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.