Abstract
The apparent resistance of a finite-thickness layer measured with a four-electrode plunge probe depends on the electrode insertion depth, electrode spacing, and layer thickness, as well as the resistivity ratio of an underlying layer. A physical model consisting of air, a saline solution layer, and an agar layer simulates the real situation of resistivity measurement. The saline layer represents the finite-thickness layer whose resistivity is to be measured by a plunge electrode probe, and the agar layer represents an underlying perturbing layer. A micropositioner controls the insertion depth of the four electrodes into the saline solution. With the apparent resistance measured on a semi-infinite-thickness layer of saline solution as standard, measurement results show decreasing apparent resistance and increasing error with increasing electrode insertion depth. This information is important for correct measurement of myocardial resistivity in vivo and in vitro.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.