Abstract
A method of using CO2 laser interferometer is described for the density measurement of intermediate region (the product of denstity n and plasma diameter d is around 1015 cm-2). Michelson type interferometer is adopted and one-tenth of fringe corresponds n x d≅1.05 x 1015 cm-2. Ge-Cu photoconductor cooled by liquid helium is used as a dector with fast time reponse. These are applied for the density measurement of plasma of BSG device8)9) and it is confirmed that this method is reliable and easy technique for the purpose.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.