Abstract

A new density comparison apparatus based on a pressure-of-flotation method (PFM) is presented. Density comparison measurements are performed for a new silicon crystal of the National Metrology Institute of Japan (NMIJ) for the density standard and determination of the Avogadro constant. The density comparison measurement of silicon crystals by the new PFM apparatus has an estimated relative standard uncertainty of 4.0 × 10−8. Adjusted values of density are evaluated from the data of the PFM and absolute measurements using a least-squares algorithm, where correlations are taken into consideration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.