Abstract

The study of human cognition, and preoperative functional brain mapping, are facilitated through the use of magnetoencephalography (MEG). However, the noise present in such recordings is significant relative to the signals of interest. To circumvent this issue multiple trials are performed for the same task and an ensemble average is performed to increase the signal-to-noise/interference ratio (SNIR). Unfortunately, large numbers of trials (100–500) are required to achieve a sufficiently large SNIR. This paper describes a simple denoising technique which employs spatial averaging to potentially reduce the number of required trials. The N trials from each of the 274 channels are first averaged. The 274 averaged channel estimates are then temporally low passed filtered, using a zero-phase filter. Finally, a spatial average is performed, where each channel is assigned the average value of a number of the most similar channels, including itself. The most similar channels are identified using correlation. The technique is applied to an auditory and somatosensory evoked response data set. A gain of approximately 1– 1.5 dB is observed over low-pass filtering alone.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.