Abstract

It is demonstrated in this report that a conventional strong-acid cation-exchange column can exhibit reversed-phase chromatographic behavior simultaneously with ion-exchange. Adjusting the pH to control cation retention has no effect on the retention of neutral organic analytes. Likewise, changes in the methanol content of the mobile phase to adjust organic analyte retention causes only a small decrease in retention of metal ions in the 0 to 10% (v/v) methanol range, and no significant effect beyond that. Linear calibration behavior of both metal cations and neutral organic analytes is found on this column over three-order of magnitude. Examples of simultaneous metal cation–neutral organic separations in both the isocratic and gradient modes are shown, with conductivity detection for the metal ions and UV for the organic analytes. An isocratic separation of metal ions and neutrals in a vitamin pill is also demonstrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.