Abstract
Simultaneous imaging of the SPECT tracer 131I and PET tracer 18F is important in the diagnosis of high- and low-grade thyroid cancers because high-grade thyroid cancers have high 18F-FDG and low 131I uptake, while low-grade thyroid cancers have high 131I and low 18F-FDG uptake. In this study, Na131I and 18F-FDG were simultaneously imaged using the Compton-PET system, in vivo. The angular resolution and sensitivity of the Compton camera with 356 keV gamma ray measured using a 133Ba point source were 12.3° and 2 × 10−5, respectively. The spatial resolution and sensitivity of PET were measured with a 22Na point source. The transaxial and axial spatial resolutions of the PET at the center of the FOV were 1.15 mm and 2.04 mm, respectively. Its sensitivity was 1.2 × 10−4. In-vivo images of the 18F and 131I isotopes were simultaneously acquired from mice. These showed that 18F-FDG was active in the heart, brown fat, and brain, while Na131I was active in the thyroid, stomach, and bladder. Artifacts were found in the Compton camera images when the activity of 131I was much lower than that of 18F. This study demonstrates the potential of simultaneous clinical imaging of 18F and 131I.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.