Abstract

Local backside wear measurements on ultra-high molecular weight polyethylene (UHMWPE) tibial inserts in LCS mobile bearing knee prostheses have been performed using a new radioisotope tracing technique. The radioisotope tracers 97Ru and 101mRh were synthesized via a fusion evaporation reaction and recoil-implanted into cylindrical plugs of UHMWPE. The labelled plugs were carefully fitted into tibial inserts at two relevant locations. With bovine serum acting as a lubricant, the tibial inserts were then worn in vitro for 500,000 and 780,000 cycles, respectively, in a pneumatic knee motion simulator. Results reflect the non-linear change of wear during the wear-in phase and its evolution to a long-term steady-state rate. This new technique shows potential for extracting localized wear rates across the backside of a tibial insert in order to develop a comprehensive backside wear model.