Abstract

ISO/TC 198 Working Group 2 is charged with the revision of ISO Standard 11137:1995. In reviewing dose setting Method 1, it has attempted to rationalize the selection of the lower limit of average bioburden for product to which Method 1 can be applied. Provisionally, this limit has been set at 1.0. The Working Group has also considered three proposals for handling sterilization dose selection for product with an average bioburden less than 1.0.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.