Abstract

Incorrect lead selection and unexpected ECG artifact during an attempted resuscitation after inhospital cardiac arrest resulted in undetected lack of cardiac monitoring for approximately 13 minutes. The patient was finally countershocked and regained a spontaneous pulse but was determined to have experienced profound neurologic damage and died shortly after being extubated. This type of failure may be common, particularly with older monitor/defibrillators. Caregivers, health care organizations, and device manufacturers should be aware of this potential problem and institute preventive measures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.