Abstract

A two-dimensional delay line readout to be used with large-diameter (127-mm) microchannel plates (MCPs) has been developed. High spatial resolutions (25 mu m FWHM for an MCP gain of 2*10/sup 7/) and excellent linearity of position readout (less than +or-65- mu m deviation from linearity) have been achieved in a test setup using 30-mm-diameter MCPs coupled to 140-mm*140-mm delay lines. Single-pass delays of only 176 ns allow system deadtimes of less than 1 mu s. This type of readout will be applied to imaging X-ray detectors to be used in protein crystallography. >

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.