Abstract

Delays in signal transmission and reception paths of an Earth station must be calibrated accurately, and their variation should be reduced in the two-way satellite time and frequency transfer (TWSTFT). We have developed a portable TWSTFT station using a modem built by the National Institute of Information and Communications Technology (NICT) and performed calibration between NICT and Telecommunication Laboratories. An uncertainty of about 1 ns in the differential calibration was achieved. In addition, we have monitored the delay variation in the Earth station using a measurement system installed at NICT to improve accuracy of the TWSTFT. The contributed instability was 50 ps/day

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.